The Thermo Orion PSTAR2146 bench-top pH/ISE meter kit comes with the 9512HPBNWP ammonia electrode, 927007MD temperature probe and 096019-WA stirrer. The Thermo Scientific Orion Pro Star PI214 pH/ion bench meter is designed to provide accurate, reliable electrochemistry testing. Featuring an intuitive interface, enhanced data reporting and robust functionality, the Orion Pro Star PI214 meter delivers advanced performance in a modern and simplified package. The Orion Pro Star PI214 meter features a wide range of parameter capabilities.
pH Range | -2.000 to 20.000 pH |
---|---|
Resolution | 0.1, 0.01 or 0.001 pH |
Relative accuracy | ± 0.002 pH |
Calibration points | 1 to 5 points |
Calibration method | Automatic buffer recognition, manual entry option |
Calibration buffer sets | USA (1.68, 4.01, 7.00, 10.01, 12.46) DIN (1.68, 4.01, 6.86, 9.18, 12.46) Pure Water Buffers (4.10, 6.97, 9.15) Standard (1.00, 3.00, 6.00, 8.00, 10.00, 13.00) Up to 5 custom buffers |
Slope type | Linear or segmented |
Calibration summary | Slope and Eo offset, option to view as point-to-point graph |
Isopotential point | 7.000, user adjustable |
Input impedance | >10^12 ohms |
mV Range | -2000.0 to 2000.0 mV |
Resolution | 0.1 mV |
Relative accuracy | ± 0.2 mV or ±0.05 % of the reading, whichever is greater Relative |
mV / ORP Range | -2000.0 to 2000.0 mV |
Resolution | 0.1 mV |
Relative accuracy | ± 0.2 mV or ±0.05 % of the reading, whichever is greater |
Calibration points | 1 point |
Calibration method | ORP or EH |
Calibration offset | ±250 mV |
Ion Range | 0.0001 to 19999 |
Resolution | 0.0001 minimum, 1 to 5 digits |
Relative accuracy | ± 0.2 mV or ±0.05 % of the reading, whichever is greater |
Units | ppt, ppm, ppb, mg/L, M, mM, %, none or user-defined |
Calibration points | 1 to 5 points |
Calibration method | Automatic standard recognition, manual entry option |
Calibration standards | 1.000, 10.00, 100.0, 1000; 10000 or up to 5 custom standards |
Slope type | Linear or segmented |
Calibration summary | Slope and Eo offset, option to view as point-to-point graph |
Isopotential point | 1.000, user adjustable |
Blank correction | Yes |
Low level stability | Yes |
Incremental techniques | Single known addition, single known subtraction, double known addition, double known subtraction, analyte addition and analyte subtraction |
Temperature Range | -30.0 to 130.0°C, -22.0 to 266.0°F |
Resolution | 0.1°C, 0.1°F |
Relative accuracy | ±0.3°C, ±0.5°F |
Calibration points | 1 or 2 points |
Offset adjustment | Up to ±5.0°C |
Temperature source | Automatic with ATC sensor, ambient with built-in meter temperature sensor or manual entry Measurement channels |
Measurement parameters | pH, mV, ORP or ion concentration with temperature |
Read types | Continuous, Auto-Read or Timed |
Timed reading intervals | 3 seconds to 2 hours |
Measurement layout | Customizable options for basic, advanced or user-defined |
Measurement stability | Automatic, fast, medium or slow |
Sample ID | Auto-incremental or manual, up to 25 alphanumeric characters |
Electrode ID | Up to 20 alphanumeric characters |
Graph / chart view | Single parameter with temperature |
Set value alarm | 1 high and 1 low value per parameter |
Last calibration details | Calibration summary with date and time |
Display | 7" high resolution color graphic LCD |
Touchscreen type | Capacitive with lab glove compatibility |
Onscreen prompts | On-demand information screens; onscreen user guide and step-by-step calibration guidance |
Time and date | Yes |
Time format | 12 hour or 24 hour |
Date format | Selectable as MM-DD-YYYY, DD-MM-YYYY or DD-MMM-YYYY |
Instrument name | Up to 20 alphanumeric characters |
Secure mode | 3 administrator and 30 user accounts with individual password protection, custom user access permissions and audit trail |
Username | Optional, up to 20 alphanumeric characters |
Methods | 10 |
Data outputs | USB flash drive, computer or compact ink printer |
Data output format | Report, CSV or print |
Data log | 10,000 data sets with time and date |
Data log export | Customizable options for basic (short), intermediate, advanced or all data |
Calibration log | 20 per parameter with time and date |
Calibration log export | Customizable options for basic (short), intermediate, advanced or all data |
Calibration reminder | 1 alarm per parameter |
Calibration alarm interval | 1 hour to 30 days or off |
Audible signal | User set for button press, notifications and alarms, stable reading |
Maintenance Reminder | Up to 5 electrode maintenance reminders and up to 5 solution maintenance reminders |
Sleep mode | Selectable time interval of 5, 10, 20, 30, 60 minutes or off |
Meter tests | Self-test and stability test |
Meter tours and guides | Calibration modes, measurement mode, meter connections, electrode stand assembly |
Stirrer probe inputs | 1 |
Stirrer probe speeds | 5 selectable speeds from 250 RPM to 3600 RPM |
Electrode inputs | 9-pin MiniDIN |
Data outputs | 1 USB-A port for USB flash drive; 1 export port for computer or compact ink printer |
Accessory input | 1 USB-A port for USB mouse, USB keyboard or USB barcode scanner |
Memory | Non-volatile, to preserve data and settings in case of power loss |
Display languages | Chinese, English, French, German, Italian, Japanese, Korean, Portuguese or Spanish |
Meter warranty | 2 years |
Certifications | CE, TUV 3-1, FCC Class A |
Enclosure | IP-54 |
Power | 100-240 VAC, 50-60Hz, 9 DC adapter, 2.0A |
Dimensions (L x W x H) | 216.3 mm x 153.2 mm x 91.62 mm, 8.52" x 6.03" x 3.61" |
Weight | 1.00 kg, 2.20 lbs. |
Includes | Orion Pro Star PI214 pH/ion bench meter ammonia kit, includes meter with stand, high-performance ammonia ion selective electrode, ATC temperature probe, stirrer probe, ammonia 1000 ppm standard, ammonia ISA, ammonia electrode storage solution |