METER ONLY, DOES NOT INCLUDE ELECTRODES
The Thermo Scientific Orion LSTAR1130 benchtop dissolved oxygen meter has a measuring range of 0.0 to 300.0% saturation with a resolution of 0.1, 1% saturation and does not come with a probe. The Thermo Scientific Orion Lab Star DO113 Dissolved Oxygen Bench Meter is ideally suited for laboratories with users of varying levels of technical expertise and who require a simple-to-calibrate and use meter for routine dissolved oxygen testing. Scientists have trusted Thermo Scientific Orion Lab Products for over 60 years to provide them with accurate and reliable analytical testing instruments in the lab and in the field. Measure dissolved oxygen as percent saturation, concentration in mg/L or concentration in ppm simultaneously with temperature using the Orion Lab Star DO113 DO Bench Meter. To enable comprehensive measurement capabilities, measure dissolved oxygen in a range of 0.0 to 300.0 % saturation, 0.00 to 30.0 mg/L or ppm and measure temperature in a range of -5.0 to 105.0 °C or 23.0 to 221.0 °F. For proficient SOP compliance, perform 100% saturation, 0% saturation and custom concentration calibrations for dissolved and one-point temperature adjustment calibrations. Featuring a large, backlit display for easy viewability and small footprint to save bench space, this dissolved oxygen meter is ideally suited for simplified, budget-friendly lab testing. Use the onscreen stabilizing and ready indicators to help identify when DO meter readings are complete. View active dissolved oxygen and temperature calibration logs with date and time stamp to ensure accurate, reproducible measurements. Record key measurements using the 500-point data log with date and time stamp. Output data logs and calibration logs to a printer or computer for long-term data storage.
Range | Saturation: 0.0 to 300.0% Concentration: 0.00 to 30.0 mg/L or ppm Temperature: -5.0 to 105.0°C (23.0 to 221.0°F) |
---|---|
Resolution | Saturation: 0.1, 1% Concentrration: 0.01, 0.1 mg/L or ppm Temperature: 0.1°C (0.1°F) |
Relative accuracy | Saturation: ±2% ±1 LSD Concentration: ±0.02 mg/L or ppm ±: 1 LSD Temperature: ±0.3°C (±0.5°F) |
Calibration | Saturation: 10% and 0% Concentration: 1 point, custom value Temperature: 1 point |
Barometric pressure compensation | 400.0 to 850.0 mmHg; 0.533 to 1.133 bar |
Salinity compensation | 0.0 to 50.0 ppt |
Temperature offset adjustment | Up to ±5°C |
Display | 5 inch backlit LCD |
Time and date | Yes |
Time format | 12 hour AM/PM or 24 hour |
Date format | Selectable, DD-MM-YYYY or MM-DD-YYYY |
Read types | Continuous, Auto-Read, Timed |
Timed interval | 5 seconds to 60 minutes |
Data log | 500 sets with time and date |
Calibration log | Active dissolved oxygen, temperature with time and date |
Calibration alarm | Yes, 1 to 168 hours or off |
Sleep mode | Yes, optional |
Self-test | Automatic with startup |
Memory | Non-volatile |
Input | 9-pin MiniDIN |
Data output | Computer or printer |
Data output format | CSV or print |
Certifications | CE, cTUVus, FCC Class B, CES-001, KC, RCM |
Enclosure | IP-54 |
Power | 100-240 VAC, 50-60Hz, 9 DC adapter, 1.3A |
Dimensions(L x W x H) | 187.2 mm x 155.2 mm x 61.7 mm, 7.37 x 6.11 x 2.43 inches |
Weight | 700 grams, 1.54 lbs. |
Environment | Indoors |
Altitude | Up to 2,000 Meters |
Operating temperature | 5°C to 45°C |
Operating relative humidity | 5 to 85%, non-condensing |
Storage temperature | -20°C to 60°C |
Storage relative humidity | 5 to 85%, non-condensing |
Mains fluctuation | ±10% of range (100-240VAC) |
Installation category | II |
Pollution degree | 2 |
Protection class | III |
Includes | Meter, electrode stand, PC cable, universal power adapter |